Performance of the eazyplex® BloodScreen GN as a simple and rapid molecular test for identification of Gram-negative bacteria from positive blood cultures

GND
1312983647
Affiliation
Institute of Medical Microbiology, Jena University Hospital
Bach, Katharina;
Affiliation
Institute of Medical Microbiology, Jena University Hospital, Friedrich Schiller University, Jena, Germany
Edel, Birgit;
Affiliation
Division of Infection Control and Infectious Diseases, RWTH Aachen University Hospital, Aachen, Germany
Höring, Steffen;
GND
1031173161
Affiliation
Institute of Medical Microbiology, Jena University Hospital
Bartoničkova, Lucie;
Affiliation
Institute of Medical Microbiology, Jena University Hospital, Friedrich Schiller University, Jena, Germany
Glöckner, Stefan;
Affiliation
Institute of Medical Microbiology, Jena University Hospital, Friedrich Schiller University, Jena, Germany
Löffler, Bettina;
Affiliation
Division of Infectious Diseases and Tropical Medicine, Department of Medicine I, Medical University of Vienna, Vienna, Austria
Bahrs, Christina;
Affiliation
Institute of Medical Microbiology, Jena University Hospital, Friedrich Schiller University, Jena, Germany
Rödel, Jürgen

The LAMP-based eazyplex® BloodScreen GN was evaluated for the detection of frequent Gram-negatives directly from positive blood culture (BC) bottles. A total of 449 BCs were analyzed. Sensitivities and specificities were 100% and 100% for Escherichia coli , 95.7% and 100% for Klebsiella pneumoniae , 100% and 100% for bla CTX-M , 100% and 100% for Klebsiella oxytoca , 100% and 99% for Proteus mirabilis , and 100% and 99.8% for Pseudomonas aeruginosa , respectively. The time to result ranged from 8 to 16 min, plus about 6 min for sample preparation. The eazyplex® BloodScreen GN is a reliable molecular assay for rapid BC testing.