000K utf8 0500 Ou 1100 $c2002 1500 ger 2051 10.1515/bmte.2002.47.s1b.566 2240 3000 Husar, Peter 3010 Berkes, Sebastian 3010 Götze, Andreas 3010 Henning, Günter 3010 Plagwitz, Kai-Uwe 4000 Verbesserung des SNR bei mehrkanaligen EEG-Ableitungen [Husar, Peter] 4060 4 Seiten 4209 The main problem in measurements of the focal VEP (Visual Evoked Potential) is its weak SNR (Signal-to-Noise Ratio). The most common methodfor enhancement of the SNR is the Stimulus synchronized averaging. For study ofsingle responses other ways in SNR improvement are needed. In this contribution a new method based on space-time selective measurement is introduced, which can be interpreted äs beaming a Signal source. Since the anatomical structures of sources generating the focal VEP are known in general and if the electrode positions are of sufficient density over the visual cortex, a source beamer can be realized by controlling the channels' delays. 4950 https://doi.org/10.1515/bmte.2002.47.s1b.566$xR$3Volltext$534 4961 https://www.db-thueringen.de/receive/dbt_mods_00040466 5051 610 5550 beam forming 5550 multichannel processing 5550 sensor array 5550 visual evoked potential